The article describes a simple, sensitive, rapid, accurate and precise spectrophotometric method for the estimation of alfuzosin hydrochloride in bulk and in pharmaceutical dosage forms. The wavelength maximum for alfuzosin hydrochloride was found to be 245 nm. The different analytical parameters were determined according to ICH Q2 (R1). Beer’s law was obeyed in the concentration range of 2-10 µg/ml. Limit of detection and limit of quantitation were found to be 0.0196 µg/ml and 0.0593 µg/ml respectively. The % recovery of drug for proposed method was 101.5760% indicating no interference of excipients. The precision of method was found within the limits.