A simple, sensitive and economic spectrophotometric method has been developed for determination of Eperisone hydrochloride in a pharmaceutical formulation. In the proposed method, the absorbances were measured at 261.0 corresponding to the absorbance maxima of Eperisone hydrochloride in distilled Water. Linearity range was observed in the concentration range of 3–24 μg/ml for Eperisone hydrochloride. Concentration of each drug was obtained by the calibration curve of absorbance vs concentration plotted. LOD and LOQ were found to be 0.214 and 0.649 respectively. The method was validated statistically and recovery study was performed to confirm the accuracy of the method. The method was found to be rapid, simple, accurate and precise.